Part Number Hot Search : 
C330K MMSZ5 IRFP2 28F16 BUV26AF C330K 39162 IRFP260P
Product Description
Full Text Search
 

To Download TM300 Datasheet File

  If you can't view the Datasheet, Please click here to try to view without PDF Reader .  
 
 


  Datasheet File OCR Text:
 TRUMPower
D E SCR IPTION
The TM300 series comprising single and multiple output models for 200 to 300 watts of continuous output power is specially designed for medical applications. They operate at 90 to 264VAC input voltage without the need of a selector strap. All auxiliary outputs are with magnetic amplifier to keep regulation. The units are constructed on a printed circuit board with a U-bracket for mechanical support and heat sinking. A cover-and-fan assembly can be added during manufacturing.
300 WATT MEDICAL POWER SUPPLIES
TM300 SERIES
( LV D )
RoHS
F EATU R E S
EN61000-3-2 class A and D compliant Pow er Factor 0.98 typical O v e r v o l ta g e p r o t e c t i o n Short-circuit protection Pow er Fail Detect (PFD) 100% burn-in at full rated load Remote sense on output #1 and output #2 Remote inhibit - TTL high disables output C o m p l i a n t w i t h R o H S r e q u i r e m e n ts
Saf et y Sta n da rd A ppr o va ls
UL 60601-1 CSA C22.2 No. 601.1
TUV EN60601-1
INPUT SPECIFICATIONS
Input voltage :
Input frequency: Input current : 90 to 264VAC 47 to 63Hz 4.7A (rms) for 115VAC 2.3A (rms) for 230VAC Earth leakage current: 100A max. @115VAC, 60Hz 220A max. @ 230VAC, 50Hz
E N V IR O N M ENTA L SP E C IF ICATI O N S
Operating temperature Storage temperature Relative humidity Derating Cooling 0C to +70C -40C to +85C 5% to 95% non-condensing Derate from 100% at +50C linearly to 50% at +70C 200/250/300 watts continuous output power at 35 CFM forced air cooling or 100/125/150 watts at convention cooling
OUTPUT SPECIFICATIONS
Output voltage/current Total output power Ripple and Noise Overvoltage protection See rating chart See rating chart 2% peak to peak maximum Provided on output #1 only; set at 115-140% of its nominal output voltage All outputs protected to short circuit conditions All outputs 0.04%/C maximum Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500s after a 25% step load change TTL logic high for normal operation and TTL logic low upon loss of input power. This signal appears at least 1ms prior to master output dropping 5% below its nominal value. This signal also provides a minimum delay of 100ms after master output is within regulation. Requires an external TTL high level signal to inhibit outputs for standard models. Switching frequency Power factor Efficiency Hold-up time Line regulation Inrush current
GENERAL SPECIFICATIONS
KHz 10KHz 0.98 typical 70% minimum on all models 12 msec minimum at 110VAC 0.2% maximum at full load 30 amps @ 115VAC or 60 amps @230VAC, at 25C cold start 4000VAC from input to output Withstand voltage 1500VAC from input to ground 500VAC from output to ground 300,000 hours minimum at full load MTBF at 25C ambient, calculated per MIL-HDBK- 217F EMC Performance (EN60601-1-2: 2001) EN55011: Class B conducted, Class B radiated EN61000-3-2: Harmonic distortion, Class A and D EN61000-3-3: Line flicker EN61000-4-2: ESD, 8KV air and 6KV contact EN61000-4-3: Radiated immunity, 3V/m EN61000-4-4: Fast transient/burst, 2KV EN61000-4-5: Surge, 1KV diff, 2KV com. EN61000-4-6: Conducted immunity, 3Vrms EN61000-4-8: Magnetic field immunity, 3A/m EN61000-4-11: Voltage dips, 30% reduction for 500ms, 60% reduction for 100ms and >95% reduction for 10ms
Overcurrent protection Temperature coefficient Transient response
PFD signal
Remote inhibit
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 Fax: 408-988-6622 email: sales@trumpower.com Website: www.TRUMPower.com
MEDICAL GRADE
OUTPUT VOLTAGE/CURRENT RATING CHART
Output # 1 Vnom. Imin. Imax. 3.3V 5.1V 12V 15V 24V 30V 48V 3.3V 5.1V 5.1V 5.1V 5.1V 5.1V 5.1V 5.1V 5.1V 24V 24V 24V 24V 24V 3.0A 60A 60A 3.0A 1.2A 25A 1.0A 20A 0.6A 12.5A 0.5A 10A 0.5A 6.3A 3.0A 35A 2.0A 35A 2.0A 35A 2.0A 35A 2.0A 35A 2.0A 35A 2.0A 35A 2.0A 35A 2.0A 35A 0.5A 6.3A 0.5A 6.3A 0.5A 6.3A 0.5A 6.3A 0.5A 6.3A
TM300 SERIES
Output # 2 Output # 3 Output # 4 Vnom. Imin. Imax. Tol. Vnom. Imin. Imax. Tol. Vnom. Imin. Imax. Maximum Output Power (5) 200W 300W 300W 300W 300W 300W 300W 250W 300W 300W 300W 300W 300W 300W 300W 300W 300W 300W 300W 300W 300W
Product No. TM300-S3.3 TM300-S5.1 TM300-S12 TM300-S15 TM300-S24 TM300-S30 TM300-S48 TM300-Q403 TM300-Q40 TM300-Q41 TM300-Q42 TM300-Q45 TM300-Q46 TM300-Q47 TM300-Q48 TM300-Q49 TM300-Q410 TM300-Q411 TM300-Q412 TM300-Q413 TM300-Q414
Tol. 3% 2% 2% 2% 2% 2% 2% 3% 2% 2% 2% 2% 2% 2% 2% 2% 2% 2% 2% 2% 2%
Tol.
5.1V 12V 15V 12V 12V 12V 24V 24V 12V 12V 12V 12V 24V 24V
2.0A 1.0A 0.8A 1.0A 1.0A 1.0A 0.5A 0.5A 1.0A 1.0A 1.0A 1.0A 0.5A 0.5A
22A 10A 8A 10A 10A 10A 5A 5A 10A 10A 10A 10A 5A 5A
2% 2% 2% 2% 2% 2% 2% 2% 2% 2% 2% 2% 2% 2%
12V 12V 15V 12V 12V 12V 12V 5.1V 5.1V 5.1V 5.1V 12V 5.1V 12V
0A 0A 0A 0A 0A 0A 0A 0A 0A 0A 0A 0A 0A 0A
4A 4A 4A 4A 4A 4A 4A 4A 4A 4A 4A 4A 4A 4A
4% 4% 4% 4% 4% 4% 4% 4% 4% 4% 4% 4% 4% 4%
12V 5.1V 24V 12V 24V 15V 12V 15V 24V 12V 24V 12V 15V 12V
0A 0A 0A 0A 0A 0A 0A 0A 0A 0A 0A 0A 0A 0A
4A 4A 2.5A 4A 2.5A 4A 4A 4A 2.5A 4A 2.5A 4A 4A 4A
4% 4% 4% 4% 4% 4% 4% 4% 4% 4% 4% 4% 4% 4%
NOTES: 1. Add suffix "C" for enclosed format with option fan control, e.g. TM300-Q45C. 2. All outputs are floating. It can be connected externally for positive or negative output. 3. Output #1 & #2 can be adjusted within 5% of their nominal voltage. 4. Output #3 & #4 can be adjusted within 15% of their nominal voltage. 5. 300 watts for "C" version with cover-and-fan assembly. 150 watts for U-bracket version without moving air (maximum current of output #1 & #2 derated to 50%), or 300 watts with 35 CFM forced air provided by user. 6. TM300-S5.1 is rated 200 watts with 35 CFM forced air cooling or 100 watts convection cooled. TM300-Q403 is rated 250 watts with 35 CFM forced air cooling (maximum current of output #1 & #2 derated to 50%) or 125 watts convection coole). 7. Single output models may be operated at no-load. At no-load, output voltage tolerance increases to 10%.
MECHANICAL SPECIFICATIONS Single Output Models
NOTES: 1. 2. 3. 4. 5. 6. 7. Dimensions shown in inch [mm] Tolerance 0.02 [0.5] maximum Input connector P1 is Dinkle DT-35-B01W-03 screws are M3, Nickel plated. Connector P4 mates with Molex housing 50-37-5103 and pins 5263. Connectors P2, P3, P5 and P6: M3*0.5 screw connections Weight: 1.20 kgs. (2.64 lbs.) approx. DC fan on P4 is rated at 12V/0.1A.
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 Fax: 408-988-6622 email: sales@trumpower.com Website: www.TRUMPower.com
MEDICAL GRADE
MECHANICAL SPECIFICATIONS Multiple Output Models
TM300 SERIES
NOTES: 1. 2. 3. 4. 5. 6. 7. 8. Dimensions shown in inch [mm] Tolerance 0.02 [0.5] maximum Input connector P1 is Dinkle DT-35-B01W-03 screws are M3, Nickel plated. Connector P4 mates with Molex housing 50-37-5103 and pins 5263. Connectors P2, P3, P5 and P6: M3*0.5 screw connections Output connector P7 mates with Molex housing 09-50-3041 and Molex 2878 series crimp terminal. Weight: 1.24 kgs. (2.73 lbs.) approx. DC fan on P4 is rated at 12V/0.1A.
PIN CHART Single Output Models
CONN PIN P1 (AC) P2 1 2 3
VO1 VO1 (VO1)
P4 P3 1
SIGNAL GROUND
Product No.
2
S (VO1)
3
S (VO1)
4
TM300-S5.1 TM300-S3.3 TM300-S12 TM300-S15 TM300-S24 TM300-S30 TM300-S48 Product No. TM300-S5.1 TM300-S12 TM300-S24 TM300-S48 CONN PIN TM300-S3.3 TM300-S15 TM300-S30
LIVE
NEUTRAL GROUND
PFD
P4 5
INHIBIT
P5 8
N.C.
P6 1 2
VOID
P7 3
VOID
6
N.C.
7
N.C.
9
0V (FAN)
10
FAN VO1 VO1
4
VOID
VOID
Multiple Output Models
Product No. TM300-Q403 TM300-Q41 TM300-Q45 TM300-Q47 TM300-Q49 TM300-Q411 TM300-Q413 CONN PIN TM300-Q40 TM300-Q42 TM300-Q46 TM300-Q48 TM300-Q410 TM300-Q412 TM300-Q414 CONN Product No. TM300-Q403 TM300-Q41 TM300-Q45 TM300-Q47 TM300-Q49 TM300-Q411 TM300-Q413 PIN 5 6 7 TM300-Q40 TM300-Q42 TM300-Q46 TM300-Q48 INHIBIT TM300-Q410 TM300-Q412 TM300-Q414 P1 (AC) 1 2 3 P2 P3 1 P4 2 3 4
SIGNAL GROUND LIVE NEUTRAL GROUND VO1 VO1 (VO1) S (VO1) S (VO1) PFD
P4 8 9 10
P5
P6 1 2
P7 3 4
N.C.
S (VO2)
S (VO2)
0V (FAN)
FAN
VO2
VO2
VO3 VO3
VO4
VO4
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 Fax: 408-988-6622 email: sales@trumpower.com Website: www.TRUMPower.com


▲Up To Search▲   

 
Price & Availability of TM300

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X